FAMILY PRO
Chem 400 pro
Gesan Chem 400 PRO is designed to meet the needs of medium and large laboratories, it is a discrete, random access, fully automated clinical chemistry analyzer with large sample plate for up to 115 samples on board, two refrigerated reagent dishes of 90 positions each to meet all laboratory needs. Precise and reliable optical measuring system, real-time system monitoring, perfect cleaning system that prevents contamination completely make it the perfect and reliable ally for working in extreme quality.
Chem 400 PRO is a floor model capable of performing 400 photometric tests per hour plus 400 ISE tests per hour, the easy-to-use multifunctional software offers flexibility for daily operation.
Remote diagnostic control allows service staff to resolve errors quickly and efficiently without an on-site visit.
TECHNICAL DATA SHEET
Overal performance | |
Equipment type | Fully automatic discrete, STAT priority |
Analysis rate | Constant speed 4000T/H or 800T/H with ISE |
Test principle | Colorimetry, turbidimetry, ISE |
Analysis method | End-point, kinetics, fixed-time, etc. support single / double wavelength and 1-4 multiple reagent item, linear and non linear calibration |
Simultaneous analysis item | 88 colorimetric items and 3 ISE items (K, Na Cl optional) |
Sample unit | |
Sample position | 115 sample positions, including 50 routine sample positions, 20 stat positions, 34 calibration positions, 8 QC positions, 3 wash soltion positions. Continuously cooling at calibration and QC positions to keep 5 15ยฐC within 24 hours. |
Sample cuveties specification | Standard cup, originalbiood tube, multi-specification |
Sample barcode | tube avaible (ร 12-16 mm x (25-10) mm |
12of5, code 128, code 39, UPC/EAN, code 93 | |
Sample | volume 2 ยตl - 35 ยตl, 0,1 ยตlstepping |
Sampling technology | Liquid level detection clot detection and collision detection |
Sample probe carryng rate | Automatic warm water cleaning. |
Carryng rates โค 0,1% | |
Automatic sample dilution | 3~170 items |
Reagent unit | |
Reagent probe | 2, with the function of liquid level detection and collision detection |
Reagent volume | 20 ยตl - 350 ยตl, 1 ยตl stapping |
Reagent position | Double reagent disk, 45x2, continuous 5-15ยฐC cooling within 24 hours, loading 70 ml and 20 ml standard reagent kit |
Reagent barcode | Code 128 |
Reagent probe carryng rate | Automatic warm water cleaning. |
Carryng rates โค 0,1% | Carryng rates โค 0,1% |
Reaction unit | |
Reaction cuvette | 120 positions opticalplasti cup, optical diameter is 6 mm |
Total volume of reaction liquid | 150 ยตl ~ 450 ยตl |
Reaction time | Random setting within 3~15 minutes |
Reaction temperature | 37ยฐC. ยฑ 0,1ยฐC |
Optical system | |
Light source | 20W / 12V halogen lamps. Life is 2000 hours |
Monocromator | Grating photometry |
Photoelectron road | After spectrophotometry |
Wavelenght | 340 nm, 380 nm, 405 nm, 450 nm, 480 nm, 505 nm, 546 nm, 570 nm, 600 nm, 660 nm, 700 nm, 800 nm, |
Detector | Photodiode led array |
OD linear range | 0 - 3,3 Abs |
Calibration and QC | |
Calibration method | 1 point linear method, 2 point linear method, multiple point linear method, non-linear method Calibration tracking Automatic description calibration K-value trends QC |
Calibration tracking | Automatic description calibration K-value trends |
QC metod | Real time QC, days QC and day QC |
Out of control processing | Alarming for out of control sample, record lost control reason |
Operating system | |
Pc operating system | Windows XP |
Analysis control software | English version graphical operating software |
Main function of software | Automatic calibration, automatic barcode scanning, item compounding test, reagent info management, serum index, whole reaction process monitoring, dirty cup memory evading, prevent cross-contamination procedure, patient information memory and association input, automatic report audit, data multiple parameter query, report format statistic and printing, reference range classification, alarming information classification, user operating right classification, auto- matic dormancy and wake, real-time online help. |
Report printing | Report format support the user-defined mode, QC and state information etc. |
PC configuration | CPU โฅ 2.2 GHz (dual core processor); Memory โฅ 1 G; Hard Disk โฅ 160 G; 17 inch LCD display; stylus, inkjet or laser printer (optional) |
System connection | TCP/IP network connection, standard |
OthersRS | 232C |
Power supply | Voltage AC 220V ยฑ 22V, 50 Hz ยฑ 1Hz, power 2KVA |
Peak water consumption | โค 25 L/H |
Reactiont disk costant temperature | Circulating water 2, after joining reagent, blending immediately |
Mixing needle | 7 stops, 11 steps by warm woter rinsing |
Reaction cuvette cleaning | Two diffluence for high and low concentration waste water |
Wastewater treatment | With the function of concentrated waste liquid level alarming |