TECHNICAL DATA SHEET
System Type | Fully Automatic, Random Access, STAT sample priority |
Throughput | Up to 300 tests / hour |
Analysis Methods | Endpoint (Bichromatic-Differantial-Sample blank), Kinetics, Fixed-Time, Turbidimetry |
Calibration Curve | Linear, Non Linear, Cubic - Spline, Multi Point, Log - Logit |
Sample Type | Serum, Plasma, Urine, CSF |
Sample Positions | Removable tray, with 60 positions - Optional 20 tubes + 20 Hitachi cups = 40 positions |
Sample Volume 2 - 300 ฮผl | 2 - 300 ฮผl |
Sample Cuvettes | Tubes of 12 -13 mm, 5 - 7 ml / cups of 0.5-1.5 ml (require a metal adapter) - optional 20 tubes of 12 -16 mm / 20 cups (Hitachi type). |
Sample Dilution | Yes - Automatic by software |
Stat Samples | Yes |
Sampling Technology | Capacitive Liquid Level Detection, needle shock sensor |
Sample Barcode | Internal barcode reader |
Reagent Positions | 2 Removable and Refrigerated Rack with 30 + 30 = 60 Reagent positions |
Reagent Volume | 5 - 350 ฮผl |
Reagent Bottles | 20 ml and 50 ml |
Pipetting Technology | Capacitive Liquid Level Detection. Needle shock sensor |
Reagent Storage | Cooling for both Reagent Rack with independent power supply |
Reagent Barcode | Hand barcode reader included |
Power Supply 1 | 00 - 240 Vac, 50 / 60 Hz, single phase with ground |
Size / Weight | 93 x 74 x 60 cm (W x D x H), 75 kg |
Reaction Cuvettes | 80 washable BIONEXยฎ cuvettes which allow approximately 20000 tests for a fully loaded rotor |
Reaction Volume | 210 - 400 ฮผl |
Reaction Time | Up to 11 min |
Reaction Temperature | 38ยฐC ยฑ 0.1ยฐC |
Optical Source | 1 halogen lamp (6 V, 10 W) with extended UV emission |
Optical Range | 0 - 3 Abs Linearity ยฑ0.5% from 0.1 to 1.5 Abs |
Optical Path | 6 mm |
Wavelength | 8 interference filters of 340, 405, 505, 546, 578, 600, 650, 700 nm, 1 free position, 1 solid position for dark reading |
Photo - Amplifier | Photoelectric detector, Signal amplifier, response range, 340 nm to 900 nm |
Diluter Syringe | 368 ฮผl capacity, long-life syringe |
Hydraulic System | 8 self-priming peristaltic pumps (life 800 hrs), 2 vacuum pumps Pinch Valve |
Wash Station | Needles: 5 dispensating, 6 aspiration, 1 clearing (7 step washing sequence for each cuvette). Less than 3 l/hour water consumption. |
Needle Wash | Internal and external with system solution |
Run / Reaction Monitor | Operation status monitor on screen |
Quality Control | Levey - Jennings Graphics - Westgard Algorithm |
Operator Interface | Included touch screen 10โ- Intel HD Graphics SO windows - 4 GB DDR3 SDRAM Intel Pentium 1.6 GHz - SSD 64GB |
Connections | Ethernet LAN port for LIS host communication, USB port for analyzer connection |
Printer | External Printer |
Environmental | 16 - 30 ยฐC, < 80% humidity |